Model Number 1320-0111
Device Problem Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/24/2021
Event Type  malfunction  
Manufacturer Narrative
Upon completion of the investigation any additional information will be communicated in a supplemental report.
 
Event Description
It was reported the distal locking was dorsal to the nail.The distal locking was performed using the freehand technique.This resulted in a 25 minute surgical delay.
 
Event Description
It was reported the distal locking was dorsal to the nail.The distal locking was performed using the freehand technique.This resulted in a 25 minute surgical delay.
 
Manufacturer Narrative
The reported event could not be confirmed, since the returned device is conforming to specifications and fully functional.Although a real root cause could not be determined the alleged event is most likely caused due to a suboptimal intra-operative procedure and has to be classified as user-customer-user error.Nevertheless, according to event description the surgery was completed using the ¿freehand technique¿, which is always an option.The device inspection revealed the following: the device returned passed the pre-operative function test as intended.Neither a proximal "mis-targeting" nor a distal "mis-targeting" could be confirmed.Potentially reduced accuracy in guidance is usually found during functional check (required per ifu).In case of any deviation it is realized prior to use.Pre-supposing that targeting accuracy for drilling was confirmed by pre-operative check it was concluded that the event was mainly based in the intra-operative procedure.Reasons for misaligned drilling are various.Potential miss-targeting can also be caused but is not limited by e.G.Loosening of the nail holding bolt during insertion of the nail.Repeated tightening of the nail holding screw prior to distal targeting / drilling is recommended.Not realized unintended loosening of the attachment knob (will lead to release of the drill sleeve).No use of drill with centre tip / unfavourable bone contour.Drilling without drill guiding sleeve.Using blunt or damaged drill.High forces applied to the target device during drilling eventually leading to unintended distortion in the system of drill, sleeve, target device and nail.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET DEVICE GAMMA3
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
MDR Report Key11858272
MDR Text Key252388854
Report Number0009610622-2021-00520
Device Sequence Number1
Product Code HSB
UDI-Device Identifier04546540716774
UDI-Public04546540716774
Combination Product (y/n)N
PMA/PMN Number
K123401
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 07/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1320-0111
Device Catalogue Number13200111
Device Lot NumberKME915095
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2021
Date Manufacturer Received07/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-