SMITH & NEPHEW, INC. SERV REPL ASCOPE A/C HD 30 DEG 4MM; ARTHROSCOPE Back to Search Results
Model Number 72202087S
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found distal tip damage, fiber damage, and a cracked distal lens.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Event or Problem Description
It was reported that, during the set-up of an unknown procedure the ascope a/c hd was cracked.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.Results of investigation have concluded that this unit had a "cracked distal lens" which makes it a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERV REPL ASCOPE A/C HD 30 DEG 4MM
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key11859970
Report Number3003604053-2021-00205
Device Sequence Number14588941
Product Code HRX
UDI-Device Identifier03596010644534
UDI-Public03596010644534
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K962075
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72202087S
Device Catalogue Number72202087S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2021
Initial Report FDA Received Date05/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-