| Model Number |
1013466-040 |
| Medical Device Problem Code |
Failure to Fold (1255)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation was unable to determine a cause for the reported difficulty.It may be possible that the balloon folds became compromised after the first inflation and removal causing difficulty during the attempt to re-fold for re-insertion; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Na.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a left av fistula.During re-insertion into the sheath, the re-wrap of the 4.0x40mm armada 18 balloon was loose, and re-insertion into the sheath was difficult.Eventually the balloon was able to be re-introduced into the sheath.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|