H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one conquest 40 pta dilatation catheter device was returned for evaluation.The returned device was noted to be bloody.Bunching was noted near the proximal end of the balloon.No other anomalies were noted.During functional testing, returned device was attempted to inflate using an inhouse presto inflation device.The balloon was noted to partially inflate and did not fully deflated.No other functional testing was done.The investigation is confirmed for the reported inflation and deflation problem, as the device was not able to inflated or deflated during the sample evaluation.The investigation is inconclusive for the reported difficult to remove due to the state of the device and no functional testing performed.The investigation is inconclusive for the reported difficult to open or remove packaging, as it cannot be reciprocated/tested in the lab environment.The pinching of the port holes noted during evaluation most likely lead to the reported inflation and deflation issues.The reported deflation issues most likely contributed to the difficulty in removing.However, the definitive root cause for the reported inflation problem, deflation problem, difficult to remove and difficult to open or remove packaging could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiry date: 12/2023), g3, h6 (device, method) h11: h6 (result, conclusion) h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.2020394-2021-01085-1.
|