|
Batch review performed on 3 may 2021: lot 2052849: (b)(4) items manufactured and released on 13-jan-2021.No anomalies found related to the problem.To date, no other similar events have been reported.Visual inspection performed by r&d project manager: after the end of the surgery, it was noticed that one of the plates of ref 03.51.10.0583, lot 2052849, detached from the instrument.The parts are not conforming to the drawing since the welding of the plate on the inner tube is not complete.The root cause of this failure might be related to a welding problem.During the visual inspection, it was noticed that the remaining plate is not completely welded to the inner tube.In particular, it looked like only half of the plate is welded, the remaining part is not.In addition, as it is possible to see from the microscope images of the inner tube, the lateral zone of the tube seems damaged.This might be related to the excessive power of the laser beam used for the welding.Clinical evaluation performed by medical affairs director: a small metal particle, few mm in dimension, is torn away from an instrument during spinal scoliosis correction.It is detected immediately after surgery through xray examination.The sutured wound is reopened and the small piece extracted.There is no clnical cause for this event.No clinical consequence should be expected as the particle has been promptly removed; the only remaining clinical damage is the reopening of the wound.
|