Model Number 115035
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during set up for a tka, the end of 2 hex driver (added for mis sizing guide a) were rounded off.There was no significant delay and the procedure was finished using a smith & nephew back up.Patient was not harmed.
 
Additional Manufacturer Narrative
H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The tip of the device is dull, rendering the device inoperable.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEX SCREWDRIVER
Common Device Name
PROSTH, KNEE, PATELLFEMTIBIAL, SEMI-CONSTR, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11862654
Report Number1020279-2021-04571
Device Sequence Number1009007
Product Code HRY
UDI-Device Identifier03596010052063
UDI-Public03596010052063
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number115035
Device Catalogue Number115035
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2021
Initial Date Received by Manufacturer 05/05/2021
Supplement Date Received by Manufacturer07/15/2021
Initial Report FDA Received Date05/21/2021
Supplement Report FDA Received Date07/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-