|
It was reported that, during a set up for a tka surgery, the jii xr tib non-spiked fix rod (case: (b)(4)) was not able to fully go through the jrny ii alignment tube (case: (b)(4)) and would get stuck.The procedure was completed without delay using a s+n back-up device.No patient injuries and no other complications were reported.
|
|
H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned device could not confirm the stated failure mode but could confirm that the device is dull and worn.The device shows signs of extensive use.A functional evaluation of the returned product confirmed the stated failure mode.The device was paired with the mating part but could not function, rendering the device inoperable.A review of complaint history did not reveal additional complaints for the listed device.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|