| Model Number |
401751 |
| Medical Device Problem Codes |
Failure to Advance (2524); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was selected for treatment of a moderately calcified lesion (99 percent stenosis degree) in the mildly tortuous mid lad.The device was advanced over guidewire through guideliner inside catheter, but did not make it to the lesion.It got caught on edge of catheter/guideliner and the stent dislodged.The stent was retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Only the stent was returned for investigation and subjected to a technical analysis.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that the stent was returned loose in a plastic pouch.The stent is still entangled with a snare and is severely deformed, i.E.Several stent struts are deformed at one end.The delivery system was not returned for analysis.Review of the production documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
|
| |
|
Search Alerts/Recalls
|