SMITH & NEPHEW, INC. LEGION PS OXIN FEM SZ6 LT; PROSTHESIS, KNEE, PATELLOFEMTIB, SEMI-CONSTRAINED, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 71421216
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/08/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, on (b)(6) 2020, 5 pieces of legion ps oxin fem sz6 lt was sent for material review board (mrb) evaluation due to handling/surface damage caused by a stoppage in the passivation machine transport system.Disposition instructions in the non-conformance indicated that the product should be returned to wet blast to be reworked.However, when reviewing the production order in sap, the order was not reworked but returned to the production floor under the original production order.This is being reported following the 21 cfr 803, a remedial action is being taken, there were no complaints for this issue.No patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION PS OXIN FEM SZ6 LT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMTIB, SEMI-CONSTRAINED, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11864553
Report Number1020279-2021-04606
Device Sequence Number1099064
Product Code JWH
UDI-Device Identifier00885556030059
UDI-Public00885556030059
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K043440
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number71421216
Device Catalogue Number71421216
Device Lot Number20HM17607
Was Device Available for Evaluation? Yes
Type of Report(Section G)Five-Day
Initial Date Received by Manufacturer 05/13/2021
Initial Report FDA Received Date05/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-