DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/29/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate the iabp and determined that there was no charging voltage from power management board and replaced it to fix the issue.
The fse performed all functional and safety checks to meet factory specifications.
Unit passed all functional and safety test per factory specifications.
The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that during a software upgrade, the cardiosave intra-aortic balloon pump (iabp) batteries would not charge.
There was no patient involvement, and no adverse event reported.
|
|
Search Alerts/Recalls
|
|
|