MAUDE Adverse Event Report: MICROVENTION, INC. AZUR-18 DETACHABLE HYDRO; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. AZUR-18 DETACHABLE HYDRO; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number 45-480305-L
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was returned to the manufacturer for evaluation.The investigation is ongoing.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with use of the device.
 
Event or Problem Description
It was reported that the embolization coil prematurely detached.No patient harm or intervention was reported.
 
Additional Manufacturer Narrative
The actual coil implant was not returned.The investigation of the returned pusher found that the monofilament experienced a tensile break based on the shape and profile of the end.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR-18 DETACHABLE HYDRO
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
MDR Report Key11867725
Report Number2032493-2021-00181
Device Sequence Number3193810
Product Code HCG
UDI-Device Identifier00810170017872
UDI-Public(01)00810170017872(11)200722(17)250630(10)2007225JB
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K090168
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Model Number45-480305-L
Device Catalogue Number45-480305
Device Lot Number2007225JB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2021
Initial Date Received by Manufacturer 04/21/2021
Supplement Date Received by Manufacturer05/17/2021
Initial Report FDA Received Date05/21/2021
Supplement Report FDA Received Date06/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-