• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number RONYX22534UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 05/18/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During a procedure an attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a mildly tortuous and moderately calcified lesion in the mid cx artery with 95% stenosis.It was noted that the vessel was an anomalous circumflex which arose from the aorta.The device was inspected with no issues noted.Negative prep was performed with no issues noted.The lesion was predilated.The device did pass through a previously-deployed stent.Resistance was encountered when advancing the device.Excessive force was used during delivery.It was reported that stent dislodgement occurred during delivery to/at lesion.The stent remained in the patient and was unable to be retrieved.It was informed that the stent dislodgement occurred due to the struts becoming entangled with the struts of a previously deployed stent.Once entangled, the removal of the delivery system resulted in the stent remaining at the ostial portion of the anomalous circumflex.The dislodged stent was dilated against the ostial/proximal wall of the artery with a coronary balloon.The patient is alive with no injury.
 
Additional Manufacturer Narrative
Additional information: annex d, annex g code added correction: device implant date added.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: the device was prepped per ifu with no issues noted.It was informed that the stent dislodgement occurred due to the struts becoming entangled with the struts of a previously deployed resolute onyx des stent.It was later reported that the previously deployed resolute onyx des had to have it's the entire segment dilated with a balloon after becoming entangled with the dislodged resolute onyx des, and has remained implanted.The physician believes that the previously deployed resolute onyx des may not have been optimally opposed to the arterial wall and the subsequently delivered stent caught on a strut.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
MDR Report Key11869263
Report Number9612164-2021-02026
Device Sequence Number16355040
Product Code NIQ
UDI-Device Identifier00643169556768
UDI-Public00643169556768
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/10/2022
Device Model NumberRONYX22534UX
Device Catalogue NumberRONYX22534UX
Device Lot Number0010171874
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/18/2021
Supplement Date Received by Manufacturer05/25/2021
06/24/2021
Initial Report FDA Received Date05/24/2021
Supplement Report FDA Received Date06/15/2021
06/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient Weight54
-
-