| Model Number |
IPN049841 |
| Medical Device Problem Code |
Insufficient Heating (1287)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#: (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
| |
|
Event or Problem Description
|
|
It was reported "the heater was plugged in, and would not heat up".No patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The sample was returned for evaluation.Functional testing was performed and the unit failed the current line and voltage test.It was found that the thermistor was damaged.Because the unit was received damaged, the complaint could not be confirmed and a root cause could not be established.
|
| |
|
Event or Problem Description
|
|
It was reported "the heater was plugged in, and would not heat up".No patient involvement reported.
|
| |
|