| Model Number |
06-3610 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/26/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding incorrect selection involving a metal head was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned clinician review: no medical records were received for review with a clinical consultant.Device history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: no further investigation for this event is possible at this time as no devices and/ or insufficient information was received by stryker orthopaedics.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
Brought pt to operating room to change fem.Head.Update 28/april/2021 (b)(4): spoke to rep.In revising the patient's hip due to a perceived limb length discrepancy, a 32+0 femoral head was revised to a 36 +10 femoral head.Post-operatively, discovery was made that the 36mm head had been implanted into a liner with inner diameter of 32mm.The patient was brought back to the operating room and the head and sleeve were exchanged for the appropriate size.
|
| |
|