This report is being filed after the review of a clinical research report for clinical safety and performance of milagro advance br interference screws in knee collateral ligament repairs.Findings are as follows: patient #4 , date of surgery: (b)(6) 2017.Medial collateral ligament repair.Milagro advance br 231819 right 22 female.Date of injury: (b)(6) 2016.Reason for injury: sports.Postoperative arthrofibrosis: yes (b)(6) 2017.Complication: patient had tibial eminence fracture.Re-operation: medial collateral ligament repair (b)(6) 2017.Arthrofibrosis; acl tear; lateral meniscus tear; knee loose bodies, mua (manipulation under anesthesia), partial lateral meniscectomy, loose body removal, lysis of adhesions, acl reconstruction.
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary :according to the information provided, it was reported that the patient had a tibial eminence fracture.No patient name or contact information was provided, therefore no follow ups can be completed at this time.Since the complaint device was not returned, we cannot determine a root cause for the reported event.If the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.Given that no lot number was provided, a manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|