MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 HUBLESS SILICONE FLAT DRAIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 HUBLESS SILICONE FLAT DRAIN Back to Search Results
Model Number 0070440
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that small broken pieces were found upon removal of lumber drain on (b)(6) 2021 which was placed on (b)(6) 2021.On (b)(6) 2021 they noted weakness and tear at base of silicone area without holes.And the third incident was about the drain tearing upon removal at the base.
 
Event or Problem Description
It was reported that small broken pieces were found upon removal of lumber drain on (b)(6) 2021 which was placed on (b)(6) 2021.On (b)(6) 2021 they noted weakness and tear at base of silicone area without holes.And the third incident was about the drain tearing upon removal at the base.
 
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.1 sample was confirmed to exhibit the reported failure.The device had not met specifications.The product caused the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging), used hubless flat silicone drain.Visual inspection of the sample noted that the white drain was torn near the transition from drain to tubing.This does not meet the specification "shaft must not be damaged or punched".A potential root cause for this failure could be ¿insufficient time or temperature to cure the drain¿.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "a.Check for fluid entering reliavac® evacuator.Lack of flow may indicate: - all exudate has been removed.- wound drain is clogged and may require irrigation and aspiration (consult physician).- auxiliary wall suction pressure is above 210mm hg.- deflated balloon: check all connections for air leak and wound tube perforations for exposure above the skin.If still deflated, replace evacuator.B.When not using auxiliary suction during surgical wound closure, several activations of the reliavac® evacuator may be required to establish suction because of: - air entering partially closed wound.- an operative air pocket.C.Insert safety pin into hole in collar to attach evacuator to patient's clothing or bed linen.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.To avoid the possibility of a hematoma due to wound evacuation, the instructions for use should be carefully followed.To avoid the possibility of drain damage or breakage: ¿ additional perforations should not be made in the drains.¿ avoid suturing through drains.¿ drains should lie flat and in line with the skin exit areas.¿ particular care should be taken to avoid any obstacles to the drain exit path.¿ drains should be checked during closure for free motion to avoid possibility of breakage.¿ drain removal should be done gently by hand.They should not be handled with pointed, toothed or sharp instruments which could cause cuts or nicks and lead to subsequent structural failure of the drain.Warning: surgical removal may be necessary if drain is difficult to remove or breaks.A.Use with single flat drain - 1.Place perforated wound drain within critical fluid collection area of wound.2.Draw drain tube through skin/stab wound incision until flat portion of drain is seated appropriately.3.Trim drain tube to desired length and attach to blue adapter.Connect other end of blue adapter to y-connector.4.Insert connecting tube in reliavac® port a, up to indicator ring.B.Use with two flat drains - 1.Place perforated portion of wound drains within critical fluid collection areas of wound.2.Draw drain tubes through skin/stab wound incision until flat portion of the drains are seated appropriately.3.Trim drain tubes to desired length.4.Cut off plug from closed arm of y-connector and attach blue adapters.5.Attach drains to blue adapters.6.Insert connecting tube in reliavac® port a, up to indicator ring.Caution: punctures or additional perforations should not be made in the silicone wound drain.C.Attaching to auxiliary suction - 1.Insert suction adapter into port b.2.During auxiliary suction, balloon will inflate and exudate will flow over balloon surface from port a to port b.3.To discontinue auxiliary suction, remove suction adapter and close port b.Caution: do not use with wall suction in excess of 210mm hg.D.To establish suction - 1.Open port b.2.Pump bulb until balloon fills container.3.Close port b.Note: hissing sound is normal and stops when maximum suction pressure is reached.Possible reflux of fluid to the patient is reduced during reliavac® evacuator reactivation by a built-in anti-reflux valve in port a.E.To empty container - 1.Open port b.2.Invert unit.3.Pump bulb to empty quickly.F.To re-establish suction - 1.Repeat step "d" above.G.To read fluid volume - 1.Open port b.2.Allow balloon to deflate.3.Read and record volume.4.To reactivate, repeat step "d" above." the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUBLESS SILICONE FLAT DRAIN
Common Device Name
HUBLESS SILICONE FLAT DRAIN
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key11872187
Report Number1018233-2021-03002
Device Sequence Number3215164
Product Code GBX
UDI-Device Identifier00801741090769
UDI-Public(01)00801741090769
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0070440
Device Catalogue Number0070440
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/30/2021
Supplement Date Received by Manufacturer11/10/2021
Initial Report FDA Received Date05/24/2021
Supplement Report FDA Received Date11/12/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-