| Catalog Number |
H93821 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified quantity of fluid transfer tube sets failed quality testing.This issue was further described as, ¿there was too much endotoxin on the lots¿.This event was observed during lab testing prior to patient use.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H4: device manufacture date: between december 14, 2020 to december 15, 2020.H10: a device was received for evaluation.A limulus amebocyte lysate (lal) fluid pathway test was performed on the device; the device met specifications.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|