| Catalog Number |
08P19-25 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
|
| |
|
Event or Problem Description
|
|
The customer stated that a falsely elevated alinity c magnesium result of 8.685 mg/dl was generated for patient sample id (b)(6) on (b)(6) 2021.Repeat testing was requested.The same sample retested at 1.927, 1.927, 1.902, 1.927, 1.923 and 1.894 mg/dl.The customer's normal range is 1.6 to 2.6 mg/dl.No adverse impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of historical quality data determined that there is normal complaint activity for lot 26090ud00.Trending review determined no adverse or non-statistical trend.Labeling is adequate and provides specific instructions to the customer regarding the current issue.Based on the available information, no product deficiency was identified.
|
| |
|
Search Alerts/Recalls
|