STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON INSERT CS #5 10MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/META
|
Back to Search Results |
|
| Model Number |
5531-G-510-E |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Loss of Range of Motion (2032)
|
| Date of Event |
04/29/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reported event an event regarding rom involving a triathlon insert was reported.The event was not confirmed.Method & results product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's left knee was revised.Surgeon reported that due to the patient not attending physical therapy, the patient developed a flexion contracture.A 5x10 cs poly was exchanged for a 5x9 cs poly.Rep said she may be able to provide the original implantation date, and otherwise confirmed no further information will be released by the hospital or surgeon.
|
| |
|
Search Alerts/Recalls
|
|
|