• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION 2.0 LACTOSORB SYSTEM 2.0 X 7 MM LACTOSORB SCREW; PLATE, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION 2.0 LACTOSORB SYSTEM 2.0 X 7 MM LACTOSORB SCREW; PLATE, BONE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bacterial Infection (1735); Purulent Discharge (1812); Swelling/ Edema (4577)
Date of Event 05/06/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2021-00283, 0001032347-2021-00284, 0001032347-2021-00285, 0001032347-2021-00286, 0001032347-2021-00287, 0001032347-2021-00288, 0001032347-2021-00289, 0001032347-2021-00290, 0001032347-2021-00291, 0001032347-2021-00292, 0001032347-2021-00293, 0001032347-2021-00294, 0001032347-2021-00295, 0001032347-2021-00296, 0001032347-2021-00297, 0001032347-2021-00298, 0001032347-2021-00299.Medical products 2.0 lactosorb system l-plate - left - regular, part# 915-2102, lot# 847100.2.0 lactosorb system l-plate - left - regular, part# 915-2102, lot# 847100.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 849260.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 849260.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 849260.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 849260.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 849260.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 849260.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 849260.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 849260.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 849280.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 849280.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 849280.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 849280.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 850040.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 850040.2.0 lactosorb system 2.0 x 7 mm lactosorb screw, part# 915-2301, lot# 850040.Report source foreign: (b)(6).
 
Event or Problem Description
It was reported the patient experienced an infection nine (9) months following fixation with resorbable implants after a maxillary osteotomy of the upper left jaw.During the resorption stage, the patients gums were swollen and pus was excreted.The surgeon is monitoring the patient and no revision is planned.It was reported that no further information is available.
 
Additional Manufacturer Narrative
This report is being submitted to update additional information in section b4, b5, d1, d4, d9, g3, g6, h1, h2, h6 and h10.It was reported that approximately 10 months postop, the patient presented with signs of infection during the resorption stage of the implant.During the investigation process a review of the sterile certifications were reviewed and found to be conforming with no applicable deviations.Devices were verified to have gone through acceptable sterilization process following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issues identified affecting implant safety or effectiveness, therefore implanted products are not identified as the source or contributing to the reported infection.Upon reassessment of the reported event based on additional information.This product did not cause or contribute to the reported event.This initial report submitted needs to be voided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.
 
Event or Problem Description
This follow-up report is being submitted to relay additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0 LACTOSORB SYSTEM 2.0 X 7 MM LACTOSORB SCREW
Common Device Name
PLATE, BONE
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
jennifer delaney
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key11872679
Report Number0001032347-2021-00300
Device Sequence Number19764095
Product Code JEY
UDI-Device Identifier00841036055028
UDI-Public(01)00841036055028(17)240731(10)850040
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K002083
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberN/A
Device Catalogue Number915-2301
Device Lot Number850040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/10/2021
Supplement Date Received by Manufacturer01/21/2022
Initial Report FDA Received Date05/24/2021
Supplement Report FDA Received Date01/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/31/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Other;
Patient Age30 YR
Patient SexMale
-
-