H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of a catheter break is confirmed but the exact cause remains unknown.One photo sample of a 4 fr groshong nxt catheter was provided for evaluation.The image shows a section of the catheter outside of patient use.A break in the catheter appears to be present; however, the characteristics of the break surface could not be closely inspected from the image.The location of the break in the catheter is could not be determined.Usage residue is visible within the tubing.Based on the photo samples provided, possible contributing factors include exposure to excessive tensile forces and material fatigue caused by repeated kinking of the catheter.Since the catheter was observed to contain a split in the tubing, the complaint is confirmed but the exact cause remains unknown.H3: evaluation findings are in section h.11.
|