As reported, the 5mm x 4cm x 80cm powerflex pro percutaneous transluminal angioplasty (pta) balloon catheter (bc) crossed the lesion and inflated to 10 atmospheric pressure (atm); however, it ruptured at 10 atm.Using a non-cordis balloon catheter, the user expanded the smart stent to finish the procedure.There was no reported patient injury.The device was opened in sterile field.The lesion was the common iliac artery.The powerflex pro was used at the left external iliac artery.The operator was a trained powerflex pro user.The device will not be returned for analysis as it was discarded.
|
|
As reported, the 5mm x 4cm x 80cm powerflex pro percutaneous transluminal angioplasty (pta) balloon catheter (bc) crossed the lesion and inflated to 10 atmospheric pressure (atm); however, it ruptured at 10 atm.Using a non-cordis balloon catheter, the user expanded the smart stent to finish the procedure.There was no reported patient injury.The device was opened in sterile field.The lesion was the common iliac artery.The powerflex pro was used at the left external iliac artery.The operator was a trained powerflex pro user.The product was not returned for analysis as it was discarded.A product history record (phr) review of lot 82195537 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics, although unknown, may have contributed to the reported event.However, with the limited amount of information provided and without the return of the device for analysis, it is difficult to draw a clinical conclusion between the device and the event reported.According to the safety information in the instructions for use, ¿if resistance is met during manipulation, determine the cause of the resistance before proceeding.Balloon pressure should not exceed the rated burst pressure.The rated burst pressure is based on the results of in vitro testing.(b)(4).Use of a pressure monitoring device is recommended to prevent over pressurization.Use only the recommended balloon inflation medium.Never use air or any gaseous medium to inflate the balloon.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
|