MAUDE Adverse Event Report: CORDIS CORPORATION POWERFLEXPRO 5MM4CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION POWERFLEXPRO 5MM4CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 4400504S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, the 5mm x 4cm x 80cm powerflex pro percutaneous transluminal angioplasty (pta) balloon catheter (bc) crossed the lesion and inflated to 10 atmospheric pressure (atm); however, it ruptured at 10 atm.Using a non-cordis balloon catheter, the user expanded the smart stent to finish the procedure.There was no reported patient injury.The device was opened in sterile field.The lesion was the common iliac artery.The powerflex pro was used at the left external iliac artery.The operator was a trained powerflex pro user.The device will not be returned for analysis as it was discarded.
 
Additional Manufacturer Narrative
As reported, the 5mm x 4cm x 80cm powerflex pro percutaneous transluminal angioplasty (pta) balloon catheter (bc) crossed the lesion and inflated to 10 atmospheric pressure (atm); however, it ruptured at 10 atm.Using a non-cordis balloon catheter, the user expanded the smart stent to finish the procedure.There was no reported patient injury.The device was opened in sterile field.The lesion was the common iliac artery.The powerflex pro was used at the left external iliac artery.The operator was a trained powerflex pro user.The product was not returned for analysis as it was discarded.A product history record (phr) review of lot 82195537 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics, although unknown, may have contributed to the reported event.However, with the limited amount of information provided and without the return of the device for analysis, it is difficult to draw a clinical conclusion between the device and the event reported.According to the safety information in the instructions for use, ¿if resistance is met during manipulation, determine the cause of the resistance before proceeding.Balloon pressure should not exceed the rated burst pressure.The rated burst pressure is based on the results of in vitro testing.(b)(4).Use of a pressure monitoring device is recommended to prevent over pressurization.Use only the recommended balloon inflation medium.Never use air or any gaseous medium to inflate the balloon.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLEXPRO 5MM4CM 80
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
MDR Report Key11876189
Report Number9616099-2021-04574
Device Sequence Number19722626
Product Code LIT
UDI-Device Identifier20705032059326
UDI-Public20705032059326
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112797
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model Number4400504S
Device Catalogue Number4400504S
Device Lot Number82195537
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/27/2021
Supplement Date Received by Manufacturer06/14/2021
Initial Report FDA Received Date05/25/2021
Supplement Report FDA Received Date06/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SMART STENT
-
-