• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH MEXICO POWERFLEXPRO 6MM4CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH MEXICO POWERFLEXPRO 6MM4CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 4400604S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As reported, a 6mm x 4cm x 80cm powerflex pro percutaneous transluminal angioplasty (pta) balloon catheter (bc) was inserted and inflated; but it ruptured at 10 atmospheric pressure (atm) during its initial inflation.Therefore, it was replaced with a new balloon catheter (6mm*4cm, non-cordis) and the procedure was completed.There was no reported patient injury.The lesion was the right common femoral artery.The lesion had severe calcification and stenosis.A contralateral approach was made with a sheath (6f, non-cordis).A guidewire (non-cordis) was inserted and crossed the lesion and an intravascular ultrasound (ivus) checked the lesion.Additional procedural details were requested but are unknown.The device was not returned for analysis as it was discarded in the hospital.A product history record (phr) review of lot 82157744 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics of severe calcification and stenosis may have contributed to the reported event, as it is known that calcium may damage balloon material.However, without return of the product for analysis, it is difficult to draw a clinical conclusion between the device and the reported event.According to the safety information in the instructions for use ¿if resistance is met during manipulation, determine the cause of the resistance before proceeding.Balloon pressure should not exceed the rated burst pressure.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence), will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over pressurization.Use only the recommended balloon inflation medium.Never use air or any gaseous medium to inflate the balloon.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Event or Problem Description
As reported, a 6mm x 4cm x 80cm powerflex pro percutaneous transluminal angioplasty (pta) balloon catheter (bc) was inserted and was inflated; but it ruptured at 10 atmospheric pressure (atm) during its initial inflation.Therefore, it was replaced with a new balloon catheter (6mm*4cm, non-cordis) and the procedure was completed.There was no reported patient injury.The lesion was the right common femoral artery.The lesion had severe calcification and stenosis.A contralateral approach was made with a sheath (6f, non-cordis).A guidewire (non-cordis) was inserted and crossed the lesion and an intravascular ultrasound (ivus) checked the lesion.The device will not be returned for analysis as it was discarded in the hospital.Additional procedural details were requested but are unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLEXPRO 6MM4CM 80
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CARDINAL HEALTH MEXICO
av prado sur 150 2d0 piso
ciudad de mexico 11000
MX  11000
MDR Report Key11878187
Report Number9616099-2021-04579
Device Sequence Number14342473
Product Code LIT
UDI-Device Identifier20705032059517
UDI-Public20705032059517
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112797
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2021
Device Model Number4400604S
Device Catalogue Number4400604S
Device Lot Number82157744
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/27/2021
Initial Report FDA Received Date05/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-