| Model Number |
PED-325-20 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that there was poor opening of a pipeline. the device was replaced, and the patient did not experience a ny injury or complications. angiographic results post procedure were normal. the devices were prepared according to the instructions for use (ifu).The patient was undergoing treatment for an unruptured, amorphous aneurysm located in the posterior communicating artery.The max diameter was 5.2mm, and the neck diameter was 3.6mm.The patient's vessel tortuosity was moderate.The landing zone was 2.2mm distal and 3.1mm proximal.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that it was unknown which section of the pipeline failed to open.The pipeline was not positioned in a vessel bend when the failure to open occurred.There were no additional steps or other devices attempted to open the pipeline.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the pipeline flex was returned for analysis.The distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.The distal hypotube and ptfe shrink tubing were found to be intact with no signs of elongation.The one end of the pipeline flex braid appeared not opened due to damaged braid.Other end of the pipeline flex shield braid was fully opened and moderately frayed.No bend found on the pusher.No damages were found with the tip coil, distal marker, re-sheathing marker or with the proximal bumper.No other anomalies were observed.Based on the returned device and customer's photos, the pipeline flex was confirmed to have failure to open as one end of pipeline flex braid was not opened due to damaged braid.In addition, the other end of the braid was found fully opened and moderately frayed.The damage to the braid on the ends of the pipeline flex braid is likely the results of the physician re-sheathing the device more than recommended two times.Possible cause includes patient tortuous anatomy.There was no non-conformance to specifications identified that led to the failure to open issue.Per our instructions for use (ifu): ¿begin to deliver the device using a combination of unsheathing the pipeline flex embolization device and pushing delivery wire simultaneously.After the distal end of the pipeline flex embolization device has successfully expanded, deploy the remainder of pipeline flex embolization device by pushing the delivery wire and/or unsheathing the pipeline flex embolization device.Resheathing and/or manipulation of the micro catheter, by locking down the delivery wire and moving both as a system, may facilitate expansion of the pipeline flex embolization device.The system is designed to allow for 2 full cycles of re-sheathing of the pipeline flex embolization device.Re-sheathing the pipeline flex embolization device more than 2 full cycles may cause damage to the distal or proximal ends of the braid.Do not use in patients in whom the angiography demonstrates the anatomy is not appropriate for endovascular treatment, due to conditions such as severe intracranial vessel tortuosity or stenosis.¿ medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|