| Model Number |
275-525 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for inspection and the cause of the problem could not be established.Review of the device history record revealed no issues during manufacturing and confirmed that the device conformed to specifications at time of release.Review of device complaint history from 2016 to 2021 identified a single additional case of broken tc inserts among a total of more than 2,700 wire cutters placed in commercial distribution during the aforementioned period.
|
| |
|
Event or Problem Description
|
|
A customer reported that the jaw of the wire cutter broke during an unspecified procedure and the fragments fell into the patient's body.There was a delay in surgery to find the fragments.All the broken pieces were removed and the surgery went well.The event occurred before (b)(6) 2021; the exact date is unknown.The doctors monitored the patient's condition post-surgery in case small invisible pieces remained in the body, potentially causing inflammation or other problems.Additional information was requested, but has not been forthcoming.
|
| |
|