• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. GRAV SET 20DP V/NV W/CKV 3SS 104-IN; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. GRAV SET 20DP V/NV W/CKV 3SS 104-IN; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 42493E
Medical Device Problem Codes Component Missing (2306); Free or Unrestricted Flow (2945)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: a complaint of a missing roller clamp was received from the customer.A sample was provided to aid in the investigation of this defect.Through visual inspection, the customer complaint was verified.The roller clamp was not assembled with the set.A device history record review for model 42493e lot number 21029646 was performed.The search showed that a total of 14,403 units in 1 lot number was built on 18feb2021.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.The root cause for this defect is an assembly error during the manufacturing of this set.This incident has been added to our database of reported incidents.
 
Event or Problem Description
It was reported that a grav set 20dp v/nv w/ckv 3ss 104-in had a missing roller clamp before use.The following was reported by the initial reporter: "there's no roller clamp it came missing".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAV SET 20DP V/NV W/CKV 3SS 104-IN
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
MDR Report Key11881108
Report Number9616066-2021-51141
Device Sequence Number1413335
Product Code FPA
UDI-Device Identifier50885403235373
UDI-Public50885403235373
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K820278
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,use
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/18/2024
Device Model Number42493E
Device Catalogue Number42493E
Device Lot Number21029646
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/05/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2021
Initial Report FDA Received Date05/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-