| Model Number |
FRED4508-PMA |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/23/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was implanted in the patient and not returned to the manufacturer for analysis.Post procedural images were provided, which confirmed the reported issue.Without the return and evaluation of the device, the investigation cannot determine if a condition existed with the device that would have caused or contributed to the event.The instructions for use identifies device migration as potential complications associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that a fred 27 stent was placed for the treatment of an unruptured anterior choroidal aneurysm.The fred stent was deployed successfully and with good neck coverage.The physician was advised to use a longer stent; however, elected to use a short one.At 6 months angiographic follow-up run, the fred was observed to be compacted back past the neck of the anurysm.The patient was discharged the same day after follow-up angiogram.The physician stated that they will retreat the aneurysm with another flow diverter.
|
| |
|