• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CORPORATION, ASHITAKA CAPIOX FX25 OXYGENATOR; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION, ASHITAKA CAPIOX FX25 OXYGENATOR; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number CX-FX25RWV
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/30/2021
Event Type  malfunction  
Manufacturer Narrative
Udi - not required for product code.Implanted date: device was not implanted.Explanted date: device was not explanted.Occupation- clinical engineer.Pma/510(k)- k130280.The actual lock adapter sample was only returned for evaluation.Visual inspection revealed no breakage or similar anomaly in the appearance.Magnifying inspection of the actual sample found no deformity or similar anomaly in the appearance.The inner diameter was measured.Compared to a current product sample, no difference was observed in the measured values.Reproductive testing was performed, and silicone was applied to a male connector of a factory-retained sampling system, a female connector was connected to it, and then a lock adapter was tightened up.As a result, the lock adapter came off.Ifu states: do not use if the package or device is damaged (e.G.Cracked) or any of the port caps are off.A review of the device history record and the product-release judgement record of the involved product code/lot# combination was conducted with no findings.It is likely that to be that some kind of lubricant was adhering to the male connector of the sampling line, and when the lock adapter was retightened, it may have get over the rib of the male connector and come off.However, with no return of the actual sampling line sample, the exact cause of the reported event cannot be definitively determined based on the available information.Terumo medical products (tmp) (importer) registration no.(b)(4) is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.(b)(4).
 
Event Description
The user facility reported that the capiox device was used pre-treatment.Dislodgement of lock adapter occurred during priming.The locking adapter on the arterial side of the sampling line came off even though no special force was applied.The procedure outcome was not reported.The patient was not harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX FX25 OXYGENATOR
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, 418
JA  418
Manufacturer (Section G)
TERUMO CORPORATION, ASHITAKA
reg. no. 9681834
150 maimaigi-cho
fujinomiya city, 418
JA   418
Manufacturer Contact
mary o'neill
reg. no. 2243441
950 elkton blvd.
elkton, MD 21921
8002837866
MDR Report Key11884168
MDR Text Key261948179
Report Number9681834-2021-00088
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071572
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue NumberCX-FX25RWV
Device Lot Number201201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/06/2021
Initial Date FDA Received05/26/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-