• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PIN LNR CONS NEUT +4 36IDX56OD; PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PIN LNR CONS NEUT +4 36IDX56OD; PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS Back to Search Results
Model Number 1218-36-656
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 05/11/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the reason for revision is dislocation, with liner and head exchange.Doi: (b)(6) 2019, dor: (b)(6) 2021, left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the material, manufacturing, inspection or sterile processing that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIN LNR CONS NEUT +4 36IDX56OD
Common Device Name
PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11886153
Report Number1818910-2021-11223
Device Sequence Number3710402
Product Code KWZ
UDI-Device Identifier10603295012122
UDI-Public10603295012122
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K043058
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1218-36-656
Device Catalogue Number121836656
Device Lot NumberJ07W78
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/11/2021
Supplement Date Received by Manufacturer06/21/2021
Initial Report FDA Received Date05/26/2021
Supplement Report FDA Received Date06/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DLT TS CER HD 12/14 36MM +5.0; PIN LNR CONS NEUT +4 36IDX56OD; DLT TS CER HD 12/14 36MM +5.0; PIN LNR CONS NEUT +4 36IDX56OD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient Weight95
-
-