Model Number 1012454-20
Medical Device Problem Codes Difficult to Remove (1528); Material Separation (1562); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaint and additional treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Event or Problem Description
This is being filed to report that after use with the nc trek balloon catheter, the distal portion of the catheter separated in the anatomy, during removal requiring medical intervention.It was reported that the procedure was performed to treat a lesion with heavy calcification in obtuse marginal coronary artery.The 4.5 x 20 mm nc trek balloon catheter was advanced to the target lesion but there was resistance due to the heavy calcification.When the nc trek balloon catheter was being removed there was resistance and the distal portion of the catheter separated in the anatomy.The separated distal portion remained on the guide wire.A snare was used to retrieve the separated portion.There was no resistance noted during removal of the balloon catheter.A new nc trek balloon catheter was advanced to complete the procedure successfully.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11886647
Report Number2024168-2021-04421
Device Sequence Number1266224
Product Code LOX
UDI-Device Identifier08717648152122
UDI-Public08717648152122
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model Number1012454-20
Device Catalogue Number1012454-20
Device Lot Number01003G1
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/05/2021
Initial Report FDA Received Date05/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient Weight88
-
-