| Model Number |
1012454-20 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Separation (1562); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaint and additional treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
This is being filed to report that after use with the nc trek balloon catheter, the distal portion of the catheter separated in the anatomy, during removal requiring medical intervention.It was reported that the procedure was performed to treat a lesion with heavy calcification in obtuse marginal coronary artery.The 4.5 x 20 mm nc trek balloon catheter was advanced to the target lesion but there was resistance due to the heavy calcification.When the nc trek balloon catheter was being removed there was resistance and the distal portion of the catheter separated in the anatomy.The separated distal portion remained on the guide wire.A snare was used to retrieve the separated portion.There was no resistance noted during removal of the balloon catheter.A new nc trek balloon catheter was advanced to complete the procedure successfully.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|