• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE PLATING SYSTEM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE PLATING SYSTEM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number LP4.5-SLF30
Medical Device Problem Codes Corroded (1131); Material Discolored (1170)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been received and the evaluation is pending.The root cause is unable to be determined at this time.A supplemental report will be provided when the evaluation has been completed.
 
Event or Problem Description
Information was received that during removal the plate was noticed to have visible corrosion and discoloration.
 
Additional Manufacturer Narrative
Device evaluation: the plate was received for visual and functional testing.Visual inspection revealed that there was corrosion on the plate, which confirmed the reported issue.Per the reported failure mode, functional testing was not applicable.The reported issue is being investigated under a capa.Device history review: a review of the device history record (dhr) indicated that returned was visually inspected and functionally tested prior to release.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE PLATING SYSTEM
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key11887873
Report Number3006179046-2021-00286
Device Sequence Number4322369
Product Code KTT
UDI-Device Identifier00887517021274
UDI-Public887517021274
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K192181
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLP4.5-SLF30
Device Lot Number0071704
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/30/2021
Supplement Date Received by Manufacturer07/30/2021
Initial Report FDA Received Date05/26/2021
Supplement Report FDA Received Date08/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age15 YR
-
-