DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/30/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the iabp, but was unable to reproduce the reported "iabp would not switch from battery to ac power" issue.
Performed all functional and safety checks to meet factory specifications.
Unit passed all functional and safety test per factory specifications.
The iabp was then released to the customer and cleared for clinical service.
The full event site name is (b)(6) campus.
|
|
Event Description
|
It was reported that during a training the cardiosave intra-aortic balloon pump (iabp) would not switch from battery to ac power.
Unit is not kicking back the power.
There was no patient involvement, and no adverse event reported.
|
|