Model Number ZCB00
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted, give date: not applicable, as lens was removed/replaced during the same procedure.If explanted, give date: not applicable, as lens was removed/replaced during the same procedure.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens (iol) was fully inserted in the left eye and removed during the same procedure due to a bent/broken haptic.Another johnson & johnson lens (same model and diopter) was implanted as a replacement.No further information is available.
 
Additional Manufacturer Narrative
Additional information: section d9.Device available for evaluation? yes returned to manufacturer: yes date returned to manufacturer: june 9, 2021 section h3.Device evaluated by manufacturer? yes device evaluation: visual inspection under magnification revealed viscoelastic residue on the optic body and haptics and that the lens was received cut into three pieces, which is consistent with a lens that was handled during removal.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing record review: based on the manufacturing records review, product evaluation, and historical complaint review, there is no indication of a product malfunction or product quality deficiency.A search of complaints revealed that no other complaints have been received for this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
MDR Report Key11888602
Report Number2648035-2021-07870
Device Sequence Number1389810
Product Code HQL
UDI-Device Identifier05050474531598
UDI-Public(01)05050474531598(17)231007
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/07/2023
Device Model NumberZCB00
Device Catalogue NumberZCB0000140
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/28/2021
Supplement Date Received by Manufacturer06/24/2021
Initial Report FDA Received Date05/26/2021
Supplement Report FDA Received Date06/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
-
-