TERUMO MEDICAL CORPORATION DESTINATION PERIPHERAL GUIDING SHEATH; INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient identifier - requested, not provided.Age & date of birth - requested, not provided.Patient sex - requested, not provided.Weight - requested, not provided.Ethnicity - requested, not provided.Race - requested, not provided.Implanted date: device was not implanted.Explanted date: device was not explanted.The actual device was not returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.A review of the device history record of the product code/lot# combination was conducted with no findings.
 
Event or Problem Description
The user facility reported that the destination tip of the guiding sheath was deformed.When the guiding sheath was inserted into the femoral artery, the guiding sheath encountered high resistance and the tip of the guiding sheath became deformed.The guiding sheath was not used and was replaced with another guiding sheath to continue the procedure.Subsequently, the procedure was successfully completed.There was no health damage to the patient.The estimated blood loss was less than 250cc's.There was no patient injury, medical/surgical intervention required.There were no other devices or equipment used with the reported product.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.The complaint cannot be confirmed for sheath catheter tip mechanical damage per the information received since the sample was not available for evaluation.Based on the information given, the exact root cause of the event cannot be determined.There is no indication that any manufacturing, design or quality system issues may have led to this event.Currently, no action is recommended since this risk evaluation is within the predetermined limits in the failure mode and effects analysis (fmea).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DESTINATION PERIPHERAL GUIDING SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
MDR Report Key11888776
Report Number1118880-2021-00115
Device Sequence Number3792515
Product Code DYB
UDI-Device Identifier00389701008851
UDI-Public00389701008851
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K172995
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2023
Device Model NumberN/A
Device Catalogue Number54-74501
Device Lot Number0000002266
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/30/2021
Supplement Date Received by Manufacturer06/25/2021
Initial Report FDA Received Date05/26/2021
Supplement Report FDA Received Date06/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-