• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY, INC. VALOR NAIL; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY, INC. VALOR NAIL; SCREW, FIXATION, BONE Back to Search Results
Model Number 415101020R
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device not returned at time of this report.If additional information becomes available, it will be provided on a supplemental report.
 
Event or Problem Description
It was reported that the internal compression screw could not be turned and removed.The surgeon removed the implant and inserted a new implant which extended the surgery time as well as tourniquet time and extra holes in the bone.No additional patient complications were reported.
 
Event or Problem Description
It was reported that the internal compression screw could not be turned and removed.The surgeon removed the implant and inserted a new implant which extended the surgery time as well as tourniquet time and extra holes in the bone.No additional patient complications were reported.
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.The operative technique was reviewed: "back off the internal compression screw using the 4.0mm hex driver", "while optional, the end cap prevents the intrusion of bone and soft-tissues into the nail, thereby making removal easier." if the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALOR NAIL
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
MDR Report Key11889290
Report Number0001043534-2021-00090
Device Sequence Number3596804
Product Code HWC
UDI-Device Identifier00840420125514
UDI-Public00840420125514
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K082320
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number415101020R
Device Catalogue Number415101020R
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/02/2021
Supplement Date Received by Manufacturer07/15/2021
Initial Report FDA Received Date05/26/2021
Supplement Report FDA Received Date08/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-