BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 2.25/26; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 364505
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.
 
Event or Problem Description
An orsiro drug-eluting stent system was selected for treatment of a moderately calcified lesion (99 percent stenosis degree) in the severely tortuous proximal lad.After pre-dilatation, during attempt to cross the lesion, it was detected that the stent got dislodged from the delivery system.The stent was retrieved.Another stent system was used afterwards.
 
Additional Manufacturer Narrative
Combination product: yes.The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that the stent is not dislodged but displaced on the balloon by about 234 mm in proximal direction on the distal shaft.Stent imprints are visible in the exposed balloon surface, indicating that the stent was initially crimped in between the two radiopaque markers.The stent is severely deformed at its distal end.The balloon is still well folded and shows no signs of inflation.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 2.25/26
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key11889616
Report Number1028232-2021-02882
Device Sequence Number14360682
Product Code NIQ
UDI-Device Identifier07640130439146
UDI-Public7640130439146
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/18/2022
Device Model Number364505
Device Catalogue NumberSEE MODEL NO.
Device Lot Number02200986
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2021
Initial Date Received by Manufacturer 05/26/2021
Supplement Date Received by Manufacturer07/26/2021
Initial Report FDA Received Date05/26/2021
Supplement Report FDA Received Date07/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-