• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. SPEEDSTITCH DEVICE; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. SPEEDSTITCH DEVICE; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number OM-7000
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/1901
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that the speedstitch device was not taking the needle.No case reported; therefore, there was no patient involvement.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).H10 h3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A visual evaluation showed the device was not returned in any original packaging.The overall appearance of the device shows signs of wear.The suture cartridge has been incorrectly installed prior to the needle being installed.A functional evaluation showed the triggers will deploy the tissue jaw tube and lock into place.The suture cartridge is installed incorrectly and cannot be removed without damaging it or possibly the device.The needle cannot be installed due to the suture cartridge having been installed before the needle, which is in the wrong order.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was confirmed and the root cause was associated with unintended use of the device.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEEDSTITCH DEVICE
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11890503
Report Number3006524618-2021-00665
Device Sequence Number1389922
Product Code GAT
UDI-Device Identifier00817470001402
UDI-Public00817470001402
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K042031
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOM-7000
Device Catalogue NumberOM-7000
Device Lot NumberH09500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/07/2021
Supplement Date Received by Manufacturer07/25/2023
Initial Report FDA Received Date05/26/2021
Supplement Report FDA Received Date07/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-