| Model Number |
VPR-GW-EL14 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Material Deformation (2976)
|
| Health Effect - Clinical Codes |
Vascular Dissection (3160); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
04/27/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Analysis of the reported device is in progress.A supplemental report will be submitted when the device analysis is completed.(b)(4).
|
| |
|
Event or Problem Description
|
|
Orbital atherectomy treatment was successfully performed via left radial access to the left popliteal (pa) and common femoral arteries without issue, and the orbital atherectomy device was removed from the patient's body.A balloon was inserted over the viperwire guide wire to treat the pa and balloon angioplasty was successfully performed.During the removal process the balloon became stuck, and the guide wire was pulled, however a kink was observed in the guide wire which prevented the balloon from being removed.The sheath, balloon and wire were pulled back into the mid aorta and access to the right groin was obtained in order to snare the wire from below, however this attempt was unsuccessful.The balloon, wire and sheath were then removed from the radial access site.Imaging was performed and confirmed a type f dissection in the axillary artery and a vessel spasm in the radial artery.Vascular surgery was consulted, and it was determined that no further treatment was required and the patient would be monitored, however the following day a coronary bypass surgery was performed.The patient was discharged in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
The guide wire was received at csi for analysis, engaged in a balloon catheter.A visual examination of the wire confirmed multiple bends and kinks along the guide wire, with the spring tip remaining intact and undamaged.At the conclusion of the device analysis investigation, the report that the guide wire was kinked was confirmed, however the dissection and vessel spasm events could not be confirmed through analysis.The root cause of the damaged guide wire was hypothesized to be due to handling of the wire during the procedure.The diamondback peripheral orbital atherectomy system instructions for use manual states ""handle the oad and guide wire carefully.A tight loop, kink, or bend in the guide wire may cause damage and system malfunction during use." the material inspection report for this guide wire lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.(b)(4).
|
| |
|
Search Alerts/Recalls
|