• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US CROSSLINK FIN GLENOID 44; GLENOID (KEELED AND 5-PEGGED) IMPLANT : SHOULDER GLENOID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US CROSSLINK FIN GLENOID 44; GLENOID (KEELED AND 5-PEGGED) IMPLANT : SHOULDER GLENOID Back to Search Results
Model Number 1137-95-026
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/13/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # :(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient was revised to address loosening of the glenoid at the cement to implant interface.Depuy cement was used.The surgeon found the keeled glenoid floating around inside the joint and removed it.Upon examining the other implants, it was determined that the patient had a functioning hemi arthroplasty so the surgeon left all the implants as they were all fixed.Doi: (b)(6) 2015.Dor: (b)(6) 2021.Left shoulder.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSLINK FIN GLENOID 44
Common Device Name
GLENOID (KEELED AND 5-PEGGED) IMPLANT : SHOULDER GLENOID
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11891558
Report Number1818910-2021-11327
Device Sequence Number13800339
Product Code KWS
UDI-Device Identifier10603295007401
UDI-Public10603295007401
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K052472
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1137-95-026
Device Catalogue Number113795026
Device Lot Number479659
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/13/2021
Supplement Date Received by Manufacturer07/02/2021
Initial Report FDA Received Date05/26/2021
Supplement Report FDA Received Date07/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CROSSLINK FIN GLENOID 44; GLOBAL UNITE HEAD 48X18 ECC; SMARTSET HV 20G
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-