• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ENCOR BIOPSY PROBE; BIOPSY INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ENCOR BIOPSY PROBE; BIOPSY INSTRUMENT Back to Search Results
Model Number ECP0112G
Medical Device Problem Codes Crack (1135); Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 09/2021).
 
Event or Problem Description
It was reported that during a biopsy procedure, the device allegedly had a foreign material.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one sealed lot sample of the encor biopsy probe was returned for evaluation.During visual evaluation, the sample packaging was opened and the probe was examined.The probe appeared clean.Foreign material was noted in the sample tissue chamber and trocar tip/aperture.A crack was noted on the distal end of the protective tubing.Skiving was noted to the inside of the protective tubing.No other anomalies were identified.Functional evaluation was not performed due to the nature of the complaint.Upon further microscopic evaluation, foreign material was noted on the tip of the aperture and aperture and flash was noted to the perimeter of the sample tray and to the port cap hole on the sample tray.Therefore, the investigation is confirmed for the reported foreign material issue in trocar tip/ aperture and in the sample tissue chamber.The investigation for the identified crack is also confirmed as the crack is visible on the distal end of the protective tubing.Additional ftir analysis has been completed on the returned material to determine the material compound identified within the sample chamber.Results indicate the material to be consistent with the tissue tray material compound, polyethylene and the material also contained small amounts of silicone.Based upon the results of the ftir analysis, it was determined the material is consistent with the sample tray material.A definitive root cause for the foreign material could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 09/2021), g3, h6 (device) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a biopsy procedure, the device allegedly had a foreign material.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one encor biopsy probe was returned for evaluation.During visual evaluation, the biopsy probe appeared to have blood residue and the probe was received aperture approximately 100% closed.The proximal retaining cap was glued to the probe.No damage was noted to the rear housing.An unknown fiber-like material was noted on the sample tissue chamber.No other anomalies were identified.Functional evaluation was not performed due to the nature of the complaint.Upon further microscopic evaluation, flash was noted to the perimeter of the sample tray and to the port cap hole on the sample tray.Slight burrs were noted to the right side of the sample tray handle.Therefore, the investigation is confirmed for the reported foreign material issue in the sample tissue chamber as the unknown fiber-like material was noted on the sample tissue chamber during visual evaluation.Additional ftir analysis has been completed on the returned material to determine the material compound identified within the sample chamber.Results indicate the material showing morphology consistent with natural textile fiber, such as cotton.A definitive root cause for the foreign material could not be determined.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event or Problem Description
It was reported that during a biopsy procedure, the device allegedly had a foreign material.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCOR BIOPSY PROBE
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
INFUS MEDICAL (THAILAND)
706 moo 4
bangpoo ind estate
samutprakarn province 10280
TH   10280
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key11891759
Report Number2020394-2021-01116
Device Sequence Number3710128
Product Code KNW
UDI-Device Identifier00801741086267
UDI-Public(01)00801741086267
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K051158
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberECP0112G
Device Catalogue NumberECP0112G
Device Lot NumberVTDW0659
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2021
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 04/30/2021
Supplement Date Received by Manufacturer09/08/2021
11/09/2021
Initial Report FDA Received Date05/27/2021
Supplement Report FDA Received Date09/09/2021
11/10/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/16/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-