• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U357566
Medical Device Problem Codes Deflation Problem (1149); Difficult to Remove (1528); Retraction Problem (1536)
Health Effect - Clinical Codes Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.Expiry date: 07/2023.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly failed to deflate and needle stick used to deflate the balloon.It was further reported that balloon allegedly had difficult to remove form sheath.Patient reported stable.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the alleged deflation issue and difficult to remove could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiry date: 07/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly failed to deflate and needle stick used to deflate the balloon.It was further reported that balloon allegedly was difficult to remove from the sheath.There was no patient injury.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly failed to deflate and needle stick used to deflate the balloon.It was further reported that balloon allegedly was difficult to remove from the sheath.There was no patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one ultraverse 035 pta dilatation catheter loaded with an unknown 7fr sheath has been received for the evaluation.On the visual evaluation, the device appeared bloody.The distal tip of the sheath was noted to be buckled with the distal end of the balloon exposed and partially deflated.No other anomalies were noted.No other functional testing was performed.Therefore the investigation for the reported difficult to remove is confirmed as the distal end of the balloon was stucked at the distal tip of the sheath.The investigation for the reported deflation problem remains inconclusive as the functional testing was not performed due to condition of the device returned.A definitive root cause for the alleged deflation issue and difficult to remove could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035 PTA CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key11892739
Report Number2020394-2021-01120
Device Sequence Number11689924
Product Code LIT
UDI-Device Identifier00801741092961
UDI-Public(01)00801741092961
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K142261
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberU357566
Device Catalogue NumberU357566
Device Lot NumberCMEV0126
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2021
Initial Date Received by Manufacturer 05/04/2021
Supplement Date Received by Manufacturer07/05/2021
10/05/2021
Initial Report FDA Received Date05/27/2021
Supplement Report FDA Received Date07/22/2021
10/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-