| Model Number |
66800042 |
| Medical Device Problem Code |
Failure to Prime (1492)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during use, a versajet exact assy, 45 degree x 8mm don´t primed.The debridement was completed without a significant delay using a smith & nephew back up device.Patient was not harmed.No other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, used in treatment, was returned for evaluation.The visual evaluation found there was no blockage.The functional evaluation found the device primed and cut as designed ,establishing no relationship between the device and the reported event.The probable causes may include kinks, obstructions or leaks in the high pressure tube.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found other related events this investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
|
| |
|