| Model Number |
66800042 |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during set up for treatment, the versajet ii exact disposable handpiece 45 degree x 8mm primed normally, but no water came out.The treatment was completed without a significant delay using the same device.Patient was not harmed.No other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, intended for use in treatment, was returned for evaluation.The visual evaluation revealed a blockage at the distal output orifice, establishing a relationship between the device and the reported event.The root cause is a blockage.A review of the manufacturing records found that there was no evidence that the product didn't meet specifications at the time of manufacture.A complaint history review found other related events.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
|
| |
|