MAUDE Adverse Event Report: ZIMMER SURGICAL SA UNIVERSAL RECIPROCATING SAW ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL SA UNIVERSAL RECIPROCATING SAW ATTACHMENT Back to Search Results
Catalog Number 89-8509-451-20
Device Problems Mechanical Problem (1384); Device-Device Incompatibility (2919)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2019
Event Type  Injury  
Manufacturer Narrative
Complaint number (b)(4).The device was not returned to the manufacturer at the date of this report.A follow-up medwatch will be sent if the product is returned or if additional information is received.
 
Event Description
It was reported that the single trigger handpiece stops driving.Moreover, the reciprocating saw was not sitting into the hand piece and the oscillating saw was not holding blade.4 devices were referenced for this event: universal modular electric/battery single trigger handpiece part number 89-8507-400-10 serial number (b)(4).Aseptic transfer kit housing part number 89-8510-440-10 with lot number 5010187.Reciprocating saw attachment part number 89-8509-451-20 serial number (b)(4).Oscillating saw attachment part number 89-8509-450-60 serial number (b)(4).The following report is relative to the reciprocating saw attachment part number 89-8509-451-20 serial number (b)(4).The event occurred during surgery.An extension of surgery of at least 60 minutes was reported.There was no additional harm or injury to patient/operator reported.
 
Event Description
It was reported that the single trigger handpiece stops driving.Moreover, the reciprocating saw was not sitting into the hand piece and the oscillating saw was not holding blade.4 devices were referenced for this event: universal modular electric/battery single trigger handpiece part number 89-8507-400-10 serial number (b)(6).Aseptic transfer kit housing part number 89-8510-440-10 with lot number 5010187.Reciprocating saw attachment part number 89-8509-451-20 serial number (b)(6).Oscillating saw attachment part number 89-8509-450-60 serial number (b)(6).The following report is relative to the reciprocating saw attachment part number 89-8509-451-20 serial number (b)(6).The event occurred during surgery.An extension of surgery of at least 60 minutes was reported.There was no additional harm or injury to patient/operator reported.
 
Manufacturer Narrative
(b)(4).After several attempts, the reciprocating saw attachment part number 89-8509-451-20 serial number (b)(6) was not returned for complaint investigation.A device history records review was performed for this product part number 89-8509-451-20 serial number (b)(6).No issue was found during the manufacturing process that could explain the defect reported.
 
Manufacturer Narrative
(b)(4).The reciprocating saw attachment part number 89-8509-451-20 serial number (b)(6) was returned for complaint investigation.Visual and functional tests were performed.Upon receipt, it was confirmed that the device had an unscrewed driven shaft and a worn locking blades.As repair, the locking blade, the driven shaft were replaced with the truck, the slide, the spacer, the mixer and the ball bearing.After repair, the device passed final tests and it was returned to the customer.A device history records review was performed for this product part number 89-8509-451-20 serial number (b)(6).No issue was found during the manufacturing process that could explain the defect reported.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the single trigger handpiece stops driving.Moreover, the reciprocating saw was not sitting into the hand piece and the oscillating saw was not holding blade.4 devices were referenced for this event: universal modular electric/battery single trigger handpiece part number 89-8507-400-10 serial number (b)(6).Aseptic transfer kit housing part number 89-8510-440-10 with lot number 5010187.Reciprocating saw attachment part number 89-8509-451-20 serial number (b)(6).Oscillating saw attachment part number 89-8509-450-60 serial number (b)(6).The following report is relative to the reciprocating saw attachment part number 89-8509-451-20 serial number (b)(6).The event occurred during surgery.An extension of surgery of at least 60 minutes was reported.There was no additional harm or injury to patient/operator reported.Extension of surgery will be no more considered as reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL RECIPROCATING SAW ATTACHMENT
Type of Device
UNIVERSAL RECIPROCATING SAW ATTACHMENT
Manufacturer (Section D)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
plan-les-ouates, geneva 1228
SZ  1228
Manufacturer (Section G)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
plan-les-ouates, geneva 1228
SZ   1228
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key11893732
MDR Text Key265834969
Report Number0008031000-2021-00014
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeNI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 01/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/27/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number89-8509-451-20
Device Lot Number5010421
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-