• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH AIMING ARM KNOB F/03.019.009 F/MULTILOC PROX HUMERAL NAIL; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH AIMING ARM KNOB F/03.019.009 F/MULTILOC PROX HUMERAL NAIL; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Catalog Number 03.019.031
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, the patient underwent the open reduction internal fixation (orif) surgery for the proximal humeral fracture with the instruments and nail in question.The surgeon checked the nail size to be used before surgery.The surgeon was planning to use a 9.5 mm nail, but in the preoperative image, the medullary cavity was unexpectedly small, so he used an 8.0 mm nail.The intramedullary nail was inserted through the guide rod, but the medullary cavity was too narrow to get inside.The surgeon used a hammer because the nail could not be inserted (sales advised against it.The hospital did not have a reamer).The nail could be inserted up to the specified position.The surgeon inserted three proximal screws and moved on to insert the distal screws, but the drill bit interfered with the nail, and it did not go in, so the surgeon removed the drill bit, tap it with a hammer, let it go through the intramedullary nail, re-attach it to the drill bit, and drill to the opposite side.After measuring, a screw was inserted.Because the drill bit interfered on the proximal side as well, the same procedure was used for penetration, measurement, and insertion.After insertion of the distal screws, the surgeon checked with images to see if it was in, but it was difficult to see, so the surgeon once took off the insertion handle and internally and externally rotated the patient arm, and the surgeon found out that the distal screws were dislodged from the intramedullary nail.The dislodged screws were removed, and the removed insertion handle was reattached.A pre-drill was performed using a set of sleeves and a 2.4 mm k-wire to be used as a lateral stop, and a 3.0 mm re-pre-drill was performed on the drilled hole.Finally, a 3.2 mm drill was performed to insert the screws.Both proximal and distal screws were inserted using the same method.The endcap was inserted last and closed the incision after confirmation.A final check was performed on the postoperative x-ray to confirm that there was no deviation of the screw, but it was found that the outer bone was partially split at the distal end of the nail.The surgery was completed within thirty (30) minutes surgical delay.The surgeon commented that soft tissue was thick, and the sleeve was not firmly extended until it came in contact with the bone (skin incision was too small).No further information is available.This report is for one (1) aiming arm knob f/03.019.009 f/multiloc prox humeral nail.This is report 7 of 20 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIMING ARM KNOB F/03.019.009 F/MULTILOC PROX HUMERAL NAIL
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key11895014
Report Number8030965-2021-04363
Device Sequence Number17043159
Product Code FZX
UDI-Device Identifier07611819432710
UDI-Public(01)07611819432710
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.019.031
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/30/2021
Initial Report FDA Received Date05/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AIM-ARM ANT F/MULTILOC HUM NAIL SYST; AIM-ARM LAT F/MULTILOC PHN; CONNECSCR CANN F/ML HUM NAIL; CONSCR F/ML HUM NAIL SYST; DRILL BIT Ø3.2 CALIBR L145 3FLUTE; DRILL BIT Ø3.2 CALIBR L145 3FLUTE; DRILL BIT Ø3.2 CALIBR L145 3FLUTE; DRILLSL 8/3.2 F/03.010.063; DRILLSL 8/3.2 F/03.010.063; INSERT-HANDLE F/ML HUM NAIL; LOCKSCR Ø4 L24 F/NAILS TAN DBLUE; LOCKSCR Ø4 L24 F/NAILS TAN DBLUE; LOCKSCR Ø4 L24 F/NAILS TAN DBLUE; PHN MULTILOC Ø8 IZQ CAN L160 TAN; PROTECT SLEEVE 12/8 L188; PROTECT SLEEVE 12/8 L188; TROCAR Ø3.2 F/03.010.064; TROCAR Ø3.2 F/03.010.064; TROCAR Ø3.2 F/03.010.064; UNK - DRILL; UNK -HAMMER/MALLET
Patient Age63 YR
-
-