BOSTON SCIENTIFIC CORPORATION JAGWIRE; ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number M00556611
Medical Device Problem Codes Break (1069); Defective Device (2588); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event was approximated to (b)(6) 2021 as no event date was reported.The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a jagwire was used in a procedure.The procedure date is unknown.During preparation, the guidewire corewire tip was broken.The procedure was completed with a new jagwire.There were no patient complications reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
 
Event or Problem Description
It was reported to boston scientific corporation that a jagwire was used in a procedure.The procedure date is unknown.During preparation, the guidewire corewire tip was broken.The procedure was completed with a new jagwire.There were no patient complications reported as a result of this event.Additional information received on june 7, 2021: during preparation when the coating on the tip was activated using saline before inserting into patient, the physician noticed the guidewire tip was kinked/bent.
 
Additional Manufacturer Narrative
Block b3 (date of event): date of event was approximated to (b)(6) 2021 as no event date was reported.Block d4, h4: the complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Block h10: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.Conclusion code d17 is being used in lieu of an adequate conclusion code for "device not returned.".
 
Additional Manufacturer Narrative
Block b3 (date of event): date of event was approximated to (b)(6) 2021 as no event date was reported.Block d4, h4: the complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Block h10: a jagwire guidewire was returned.Visual analysis revealed that the distal tip section was bent.The ptfe area was kinked and has indentation marks.The complaint was not confirmed.Based on the condition of the returned device, engineers determined that the failure mode has been generated by the manipulation of the device.In addition, the indentation marks could have been generated during the process.Boston scientific has determined the most probable cause of this complaint is adverse event related to procedure.It is most likely that the adverse event occurred during the procedure and the device had no influence on the event which led to the reported event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported to boston scientific corporation that a jagwire was used in a procedure.The procedure date is unknown.During preparation, the guidewire corewire tip was broken.The procedure was completed with a new jagwire.There were no patient complications reported as a result of this event.Additional information received on june 7, 2021: during preparation when the coating on the tip was activated using saline before inserting into patient, the physician noticed the guidewire tip was kinked/bent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAGWIRE
Common Device Name
ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key11895197
Report Number3005099803-2021-02527
Device Sequence Number3685958
Product Code OCY
UDI-Device Identifier08714729172161
UDI-Public08714729172161
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K141820
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00556611
Device Catalogue Number5661
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2021
Initial Date Received by Manufacturer 05/07/2021
Supplement Date Received by Manufacturer06/07/2021
07/16/2021
Initial Report FDA Received Date05/27/2021
Supplement Report FDA Received Date06/24/2021
08/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-