• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US VAPR 3.5MM SIDE STR -EA; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US VAPR 3.5MM SIDE STR -EA; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE Back to Search Results
Model Number 225301
Medical Device Problem Code Blocked Connection (2888)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Udi: (b)(4).The expiration date is currently unavailable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported by the affiliate in (b)(6) that during an unknown procedure on (b)(6) 2021, it was observed that the vapr 3.5mm side str electrode device was not recognized by the main unit when it was attached to it.Another like device was used to complete the procedure without delay.There were no adverse patient consequences reported.No additional information was provided.The devices were brand new and its first use when the issue occurred.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number (u2007009), and no non-conformance was identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAPR 3.5MM SIDE STR -EA
Common Device Name
ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key11896039
Report Number1221934-2021-01687
Device Sequence Number19760453
Product Code GEI
UDI-Device Identifier10886705009251
UDI-Public10886705009251
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K082643
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number225301
Device Catalogue Number225301
Device Lot NumberU2007009
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2021
Supplement Date Received by Manufacturer06/10/2021
Initial Report FDA Received Date05/27/2021
Supplement Report FDA Received Date06/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-