| Model Number |
AVA14 |
| Medical Device Problem Codes |
Overheating of Device (1437); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Report confirmed.The device was tested and the temperature was measured to be 150.2°f.The rate of temperature rise was above threshold at 12.1°f/sec.The likely cause was identified as corroded internal tube bearing.The device user manual warnings section includes instructions that heavy side loads and/or long operating times may cause the device to overheat.If the device overheats, discontinue use and rest the device by using it intermittently, or wrap the device in a moist sterile towel.Use of an overheated device may cause injury to the patient or operator.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Repair request initiated for device with the report of overheating and will not rotate/frozen.It was reported that there was no patient impact.Repair request escalated to a product event based on reason for return.
|
| |
|