| Model Number |
FRED2513-PMA |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Codes |
Obstruction/Occlusion (2422); Unspecified Tissue Injury (4559)
|
| Date of Event |
04/23/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was implanted in the patient and not returned to the manufacturer for evaluation.Procedural or post-procedural images were not provided for review; therefore, the alleged product issue cannot be confirmed.A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The instructions for use (ifu) identifies neurological deficit as a potential complication associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that a fred 21 stent was successfully deployed in the anterior cerebral artery and the aneurysm had stagnant flow.Eight days post procedure, cta showed that the vessel was not filling, the stent had compressed, and the patient had infarction in the region of treatment.The patient is now in rehabilitation and the physician predicts their condition should improve.
|
| |
|