• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD AIRVO 2 HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD AIRVO 2 HUMIDIFIER Back to Search Results
Model Number PT101
Device Problem Loss of Power (1475)
Patient Problem Low Oxygen Saturation (2477)
Event Date 04/29/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).The pt101 airvo2 humidifier user manual states that the "airvo is for the treatment of spontaneously breathing patients who would benefit from receiving high-flow, warmed and humidified respiratory gases" and "use continuous oxygen monitoring on patients who would desaturate significantly in the event of disruption to their oxygen supply".The airvo2 user manual also states following warnings: the unit is not intended for life support.Appropriate patient monitoring must be used at all times.Loss of therapy will occur if power is lost.Prior to each patient use, ensure that the auditory alarm signal is audible by conducting the alarm system functionality check described in the alarms section.If either alarm signal is absent, do not use the unit.Contact your fisher & paykel healthcare representative.The airvo2 user manual states that "when the unit has been disconnected from the mains/utility power socket the auditory alarm will sound for at least 120 seconds.If power is reconnected in this time, the unit will automatically restart".
 
Event Description
A healthcare facility in (b)(6) reported via a fisher & paykel healthcare (f&p) field representative that a patient deteriorated following a power outage at the hospital, while using the pt101 airvo2 humidifier.There was no reported device malfunction.No further patient consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRVO 2 HUMIDIFIER
Type of Device
AIRVO 2 HUMIDIFIER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
MDR Report Key11898813
MDR Text Key253117843
Report Number9611451-2021-00602
Device Sequence Number1
Product Code BTT
UDI-Device Identifier09420012422286
UDI-Public(01)09420012422286(10)2100521189(11)180703
Combination Product (y/n)N
PMA/PMN Number
K131895
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 04/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/27/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberPT101
Device Catalogue NumberPT101
Device Lot Number2100521189
Was Device Available for Evaluation? No
Date Manufacturer Received04/30/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/03/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-